Calibration is a critical step in ensuring the accuracy and consistency of your DMIx SamplR scans. The calibration process adjusts the scanner and iPhone to match the environmental and device-specific conditions, enabling precise color and texture reproduction. It is mandatory before the first use and re-calibration is recommended on a regular basis.
When to Calibrate
- Initial Setup: Calibration is required when the SamplR is used for the first time.
- After Moving: If the scanner is relocated to a new environment.
- Device Changes: After replacing the iPhone or the phone adapter.
- Periodic Use: Recalibrate after approximately every 3 months or when scans show inconsistencies.
Required Tools
Black & White Level Calibration Targets:
Provided with the scanner. You will find them in the box containing the accessories. Please note that both calibration targets are protected by an additional piece of cardboard. Please remove this first to access both targets.
Replace the calibration paper if it becomes bent, dirty, or damaged.
Calibration Paper needs to be stored in a dark environment at any time.
Calibration Process | |
|---|---|
| 1 | Prepare the Scanner:
|
| 2 | Navigate to the settings menu in the DMIx SamplR App (wrench icon). |
| 3 | Initiate the calibration process by clicking on “Calibration State”. |
| 4 | Start the calibration process by clicking on “Next”. |
| 5 | Unpack the White Calibration Target and place it carefully inside the scanner.
|
| 6 | Close the visor, click "Continue" and wait until the progress bar reaches 100%. |
| 7 | As soon as you see "Step 2: Black Level" on the iPhone screen, open the visor and remove the white target. Place it back in the protective sleeve that came with it, which you had previously taken it out of. |
| 8 | Unpack the Black Calibration Target and place it carefully inside the scanner.
|
| 9 | Close the visor, click "Continue" and wait until the progress bar reaches 100%. |
| 10 | As soon as you see "Step 3: Transparency" on the iPhone screen, open the visor and remove the black target. Place it back in the protective sleeve that came with it, which you had previously taken it out of. |
| 11 | Close the visor, click "Continue" and wait until the progress bar reaches 100%. |
| 12 | As soon as you see "Calibration Succedded" you can click “Finish”. |
| The DMIx SamplR is now ready to be used |